Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
Chapter 1
125
GPIB Command Dictionary
CDANalyzer subsystem
:MODE <string>
:MODE?
These commands set/query the measurement state of the code domain 
analyzer.
CDAN:MODE can be overridden with the TRIG:MODE:RETR REP and 
TRIG:MODE:RETR SING trigger commands. When the trigger 
commands are used, the state of the field on the display is not updated. 
(For instance, the screen may show 
Single
 selected when 
TRIG:MODE:RETR REP is used.)
Syntax
CDAN:MODE ‘Single’
 ‘Cont’
CDAN:MODE?
Screen/field equivalent
CDAN:MODE controls the 
Single/Cont
 subfield of the 
Measurement
 
field on the 
Main
 menu of the CODE DOM screen, when the CDMA 
standard is set to “IS-95 Only”.
:PATH:GAIN <real number>
:PATH:GAIN?
These commands set/query the gain into the code domain analyzer. This 
command fine tunes the gain setting. This value is only selectable 
through GPIB. 
See also, CAN:PATH:GAIN on page  106.
Syntax
CDAN:PATH:GAIN <real number> !values from 0 to 36 dB in 0.1 
dB steps
CDAN:PATH:GAIN?
Screen/field equivalent
CDAN:PATH:GAIN has no equivalent field. This applies when the 
CDMA standard is set to “IS-95 Only”.