128
Chapter 1
GPIB Command Dictionary
CDANalyzer subsystem
:PUNit <string>
:PUNit?
:PUNit?
These commands set/query whether power measurements are absolute
power measurements or relative power measurements.
power measurements or relative power measurements.
• ‘Abs’ - Absolute code domain power displays the power in each of the
64 Walsh channels, relative to the total power inside a 1.23 MHz
bandwidth centered at the tune frequency. The 64 Walsh channels (0
through 63) are represented by a vertical bar on the analyzer’s
display.
bandwidth centered at the tune frequency. The 64 Walsh channels (0
through 63) are represented by a vertical bar on the analyzer’s
display.
• ‘Rel’ - Relative code domain power displays the power in each of the
64 Walsh channels, relative to the pilot’s power. (Pilot power is
approximately two-thirds of the total power.) The 64 Walsh codes (0
through 63) are represented by a vertical bar on the analyzer’s
display.
approximately two-thirds of the total power.) The 64 Walsh codes (0
through 63) are represented by a vertical bar on the analyzer’s
display.
Syntax
CDAN:PUN ‘Abs’
‘Rel’
CDAN:PUN?
Screen/field equivalent
CDAN:PUN controls the
CD pwr unit
field on the CODE DOM screen
when the CDMA standard is set to “IS-95 Only”.
:SAMPle:TIME <real number>
:SAMPle:TIME?
:SAMPle:TIME?
These commands set/query the time interval for the code domain
measurements.
measurements.
Syntax
CDAN:SAMP:TIME <real number> !values 0.00025 to 0.00125 s
CDAN:SAMP:TIME?
Screen/field equivalent
CDAN:SAMP:TIME controls the
Meas Intvl
field on the
Aux
menu of
the CODE DOM screen when the CDMA standard is set to “IS-95
Only”.
Only”.