194
Chapter 1
 
GPIB Command Dictionary
DECoder subsystem
DECoder subsystem
:ARM:MODE <string>
:ARM:MODE?
These commands set/query the measurement arming mode. This 
command specifies how long you want the analyzer to decode incoming 
signals.
• ‘Single’ tells the analyzer to display the information received during 
one gate time. Measurements are triggered using the   
DGAN:TRIG:ARM and retriggered using the same command. To 
disarm the measurement, use the DEC:STOP command.
• ‘Cont’ is used to automatically re-arm the analyzer and display the 
measurements continuously until the DEC:ARM:MODE ‘Single’ 
command is sent. Each measurement is overwritten by subsequent 
measurements.
Syntax
DEC:ARM:MODE ‘Single’
‘Cont’
DEC:ARM:MODE?
Screen/field equivalent
DEC:ARM:MODE controls the 
Single/Cont
 field on the SIGNALING 
DECODER screens.