GPIB Command Dictionary
DECoder subsystem
:LEVel:VOLTs <real number>
:LEVel:VOLTs?
These commands set/query the expected data signal level (after 
de-emphasis, if used). For more information about the effects of 
de-emphasis, see the “De-emphasis and Input Level” in the Reference 
Guide
.
The higher the level of signal expected by the analyzer, the higher the 
trigger level is set. The input level should be set high enough to prevent 
false triggering, but low enough to allow triggering for valid signals. 
This may require you to set the input level well below the expected level.
This command requires one of the following settings for the AF 
analyzer’s input (AFAN:INP), ‘Audio In’, ‘Audio Out’, ‘Ext Mod’ or ‘SSB 
Demod’.
This command utilizes the “Real Number Setting Syntax” on page  299, 
but does not use the :STATe command.
Syntax
DEC:LEV:VOLT <real number>
DEC:LEV:VOLT?
Screen/field equivalent
DEC:LEV:VOLT controls the 
Input Level
 field on the SIGNALING 
DECODER screen when the 
AF Anl In
 field is set to 
Audio In
Audio 
Out
Ext Mod
, or 
SSB Demod
.
:MODE <string>
:MODE?
These commands set/query the mode of the signaling decoder.
Syntax
DEC:MODE ‘Func Gen’
‘DTMF’
‘AMPS-TACS’
‘NAMP-NTAC’
DEC:MODE?
Screen/field equivalent
DEC:MODE controls the 
Mode
 field on the SIGNALING DECODER 
screen.