Chapter 1
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:ERHo?
IS-95 Only mode. This command queries the estimated rho
measurement. Estimated rho is an approximation of rho that is
calculated without taking the system out of service.
measurement. Estimated rho is an approximation of rho that is
calculated without taking the system out of service.
Syntax
MEAS:CDAN:ERH?
Screen/field equivalent
MEAS:CDAN:ERH? queries the
Est Rho
field CODE DOM screen
when the CDMA standard is set to “IS-95 Only”.
:CDANalyzer:MARKer:APOWer <meas cmnd>
:CDANalyzer:MARKer:APOWer?
:CDANalyzer:MARKer:APOWer?
IS-95 Only mode. These commands set/query the absolute power
measurement.
measurement.
This measurement utilizes the Number Measurement Syntax, but does
not use the :METer command.
not use the :METer command.
Syntax
MEAS:CDAN:MARK:APOW <meas cmd>
MEAS:CDAN:MARK:APOW?
Example
MEAS:CDAN:MARK:APOW?
Screen/field equivalent
MEAS:CDAN:MARK:APOW? reads the
Lvl
field on the
Marker
menu
of the CODE DOM screen when power (or fast power) measurement is
selected and the
selected and the
Ch pwr unit
field on the
CD Setup
menu is set to
Abs
.
This applies when the CDMA standard is set to “IS-95 Only”.