Chapter 1
 
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:ERHo?
IS-95 Only mode. This command queries the estimated rho 
measurement. Estimated rho is an approximation of rho that is 
calculated without taking the system out of service.
Syntax
MEAS:CDAN:ERH?
Screen/field equivalent
MEAS:CDAN:ERH? queries the 
Est Rho
 field CODE DOM screen 
when the CDMA standard is set to “IS-95 Only”.
:CDANalyzer:MARKer:APOWer <meas cmnd>
:CDANalyzer:MARKer:APOWer?
IS-95 Only mode. These commands set/query the absolute power 
measurement.
This measurement utilizes the Number Measurement Syntax, but does 
not use the :METer command.
Syntax
MEAS:CDAN:MARK:APOW <meas cmd>
MEAS:CDAN:MARK:APOW?
Example
MEAS:CDAN:MARK:APOW? 
Screen/field equivalent
MEAS:CDAN:MARK:APOW? reads the 
Lvl
 field on the 
Marker
 menu 
of the CODE DOM screen when power (or fast power) measurement is 
selected and the 
Ch pwr unit
 field on the 
CD Setup
 menu is set to 
Abs
This applies when the CDMA standard is set to “IS-95 Only”.