Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
Chapter 1
241
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:RHO[:ESTimated]?
IS-2000 only. This query returns the Estimated Rho when present (not 
available in FAST Power Measurement). The returned value is a 
floating point number. Estimated rho is an approximation of rho that is 
calculated when there are more signals than a pilot alone.
Syntax
MEAS:CDAN:IS2000:RHO[:EST]?
Screen/field equivalent
MEAS:CDAN:IS2000:RHO:EST? reads the 
Est Rho
 field on the CODE 
DOMAIN screen when the CDMA standard is set to “IS-2000”.
:CDANalyzer:IS2000:TIME:OFFSet?
IS-2000 only. This query returns the time offset value of the current 
Code Domain measurement as a floating point number.
Syntax
MEAS:CDAN:IS2000:TIME:OFFS?
Screen/field equivalent
MEAS:CDAN:IS2000:TIME:OFFS? reads the 
Time Ofs
 measurement 
field on the CODE DOMAIN screen when the CDMA standard is set to 
“IS-2000”.