Agilent Technologies E6380-90073 Manuale Utente
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
242
Chapter 1
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:ABSolute:POWer[:BASic][:VALue]?
IS-2000 only. This query returns an array of 128 values representing
the true (absolute) power of the Walsh channel on which the Marker is
currently positioned.
the true (absolute) power of the Walsh channel on which the Marker is
currently positioned.
NOTE
The value returned represents the absolute power of the Walsh channel
not the supplemental channel. The power of the supplemental channel
must be read directly from the
not the supplemental channel. The power of the supplemental channel
must be read directly from the
Marker Lvl
value on the CODE
DOMAIN screen, or with the
MEAS:CDAN:IS2000:TRAC:ABS:POW[:BAS]:ALL?
query.
Syntax
MEAS:CDAN:IS2000:TRAC:ABS:POW[:BAS][:VAL]?
Screen/field equivalent
MEAS:CDAN:IS2000:TRAC:ABS:POW[:BAS][:VAL]? returns values
that match the values displayed for Power, Fast Power, and Power &
Noise measurements when
that match the values displayed for Power, Fast Power, and Power &
Noise measurements when
CD pwr unit
is set to
Abs
from the
Reference
image on the CODE DOMAIN screen when the CDMA
standard is set to “IS-2000”.
:CDANalyzer:IS2000:TRACe:ABSolute:POWer[:BASic]:ACTive?
IS-2000 only. This query returns the active channel information for the
Code Domain trace when making a Power, Fast Power, or Fast Power
Synchronize measurement. The returned entity consists of an integer
representing the number of active channels followed by a four-value
record for each channel. The four values for each record are:
Code Domain trace when making a Power, Fast Power, or Fast Power
Synchronize measurement. The returned entity consists of an integer
representing the number of active channels followed by a four-value
record for each channel. The four values for each record are:
• Channel number
• Spread factor for this channel
• Power for this channel
• A fourth (ignore it) value
Syntax
MEAS:CDAN:IS2000:TRAC:ABS:POW[:BAS]:ACT?
Screen/field equivalent
MEAS:CDAN:IS2000:TRAC:ABS:POW[:BAS]:ACT? has no equivalent
field on the CODE DOMAIN screen. The information can only be
returned when the CDMA standard is set to “IS-2000”, the
measurement selected is
field on the CODE DOMAIN screen. The information can only be
returned when the CDMA standard is set to “IS-2000”, the
measurement selected is
Power
,
Fast Power
or
FstPwr Sync
, and
CD
pwr unit
is set to
Abs
on the Reference image of the CODE DOMAIN
screen.