Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
244
Chapter 1
 
GPIB Command Dictionary
MEASure subsystem
:
CDANalyzer:IS2000:TRACe:ABSolute:POWer:COMPlex:I:ACTive?
:CDANalyzer:IS2000:TRACe:ABSolute:POWer:COMPlex:COMBined:ACTive?
:CDANalyzer:IS2000:TRACe:ABSolute:POWer:COMPlex:QACTive?
:CDANalyzer:IS2000:TRACe:ABSolute:POWer:COMPlex:IS2000:ACTive?
IS-2000 only. These queries return the active channel information for 
the Code Domain trace when making a Complex Power measurement. 
The return values consist of an integer representing the number of 
active channels followed by a four value record for each channel. The 
four values for each record are channel number, spread factor for this 
channel, I value for this channel and the Q value for this channel.
Syntax
MEAS:CDAN:IS2000:TRAC:ABS:POW:COMP:I:ACT?
MEAS:CDAN:IS2000:TRAC:ABS:POW:COMP:COMB:ACT?
MEAS:CDAN:IS2000:TRAC:ABS:POW:COMP:Q:ACT?
MEAS:CDAN:IS2000:TRAC:ABS:POW:COMP:IS2000:ACT?
Screen/field equivalent
There are no fields on the CODE DOMAIN screen equivalent to the 
return value of these queries. These values are only accessible when the 
CDMA standard is set to “IS-2000”, the measurement selected is 
Complex Power
, and 
CD pwr unit
 is set to 
Abs
 on the Reference image 
of the CODE DOMAIN screen.