Chapter 1
 
GPIB Command Dictionary
RFGenerator subsystem
:FREQuency <real number>
:FREQuency?
These commands set/query the frequency of the RF generator. The Test 
Set must first be in frequency tuning mode. (See CONF:RFD ‘Freq’ on 
page  164)
This command utilizes the “Real Number Setting Syntax” on page  299, 
but does not use the :STATe command.
Syntax
RFG:FREQ <real number>
RFG:FREQ?
Screen/field equivalent
RFG:FREQ controls the 
RF Gen Freq
 field on the RF GENERATOR 
screen, when in frequency tuning mode. This field is also displayed on 
the spectrum analyzer’s 
RF Gen
 menu.
:MODulation:AOUT <string>
:MODulation:AOUT?
These commands set/query the coupling between the demodulated 
audio and the AUDIO OUT port. 
Syntax
RFG:MOD:AOUT ‘AC’
‘DC’
RFG:MOD:AOUT?
Screen/field equivalent
RFG:MOD:AOUT controls the 
Audio Out
 field on the RF GENERATOR 
screen.