Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
312
Chapter 1
 
GPIB Command Dictionary
RFGenerator subsystem
:MODulation:EXTernal:FM <real number>
:MODulation:EXTernal:FM?
These commands set/query the FM deviation of the RF generator when 
FM is applied through the modulation input port. The modulation input 
must be set to FM.
Syntax
:MOD:EXT:FM <real number>
:MOD:EXT:FM?
Screen/field equivalent
RFG:MOD:EXT:FM controls the lower subfield of the 
Mod In To
 field 
on the RF GENERATOR screen when 
FM (/Vpk)
 is selected in the 
upper subfield.
:OUTPut <string>
:OUTPut?
These commands set/query the RF output port.
Syntax
RFG:OUTP ‘RF Out’
‘Dupl’
RFG:OUTP?
Screen/field equivalent
RFG:OUTP controls the 
Output Port
 field on the RF GENERATOR 
screen. This field is also displayed on the spectrum analyzer’s 
RF Gen
 
menu when the spectrum analyzer is in fixed mode.