Chapter 1
 
GPIB Command Dictionary
SANalyzer subsystem
:MASK:FIXed:UPPer:POINts:EDIT
:MASK:FIXed:UPPer:POINts:EDIT?
This command selects/queries the point number to edit on the upper 
spectrum analyzer mask. The maximum number is 15. 
Syntax
SAN:MASK:FIX:UPP:POIN:EDIT <integer 1-15>
SAN:MASK:FIX:UPP:POIN:EDIT?
Screen/field equivalent
SAN:MASK:FIX:UPP:POIN:EDIT controls the 
EditPt
 field on the 
Mask
 
menu of the SPEC ANL screen when the 
Edit Mask
 field is set to 
Upper
, and the 
Mask Type
 field is set to 
Fix
.
:MASK:FIXed:UPPer:POINts:LEVEL1 (through LEVEL15)
:MASK:FIXed:UPPer:POINts:LEVEL1? (through LEVEL15)
These commands set/query the level of the defined points on the upper 
spectrum analyzer mask. 
Syntax
To define the levels in a four-point mask:
SAN:MASK:FIX:UPP:POIN:LEVEL1 <real number>
SAN:MASK:FIX:UPP:POIN:LEVEL2 <real number>
SAN MASK:FIX:UPP:POIN:LEVEL3 <real number>
SAN:MASK:FIX:UPP:POIN:LEVEL4 <real number>
SAN:MASK:FIX:UPP:POIN:LEVEL1?
Screen/field equivalent
SAN:MASK:FIX:UPP:POIN:LEVEL1 controls the 
Level
 field on the 
Mask
 menu of the SPEC ANL screen when the 
EditPt
 field is set to 
1
the 
Edit Mask
 field is set to 
Upper
, and the 
Mask Type
 field is set to 
Fix
.