Chapter 1
GPIB Command Dictionary
SANalyzer subsystem
:MASK:RELative:LOWer:POINts:EDIT
:MASK:RELative:LOWer:POINts:EDIT?
:MASK:RELative:LOWer:POINts:EDIT?
This command selects/queries the point number to edit on the lower
spectrum analyzer mask. The maximum number is 15.
spectrum analyzer mask. The maximum number is 15.
Syntax
SAN:MASK:REL:LOW:POIN:EDIT <integer 1-15>
SAN:MASK:REL:LOW:POIN:EDIT?
Screen/field equivalent
SAN:MASK:REL:LOW:POIN:EDIT controls the
EditPt
field on the
Mask
menu of the SPEC ANL screen when the
Edit Mask
field is set to
Lower
, and the
Mask Type
field is set to
Rel
.
:MASK:RELative:LOWer:POINts:LEVEL1 (through LEVEL15)
:MASK:RELative:LOWer:POINts:LEVEL1? (through LEVEL15)
:MASK:RELative:LOWer:POINts:LEVEL1? (through LEVEL15)
These commands set/query the level of the defined points on the lower
spectrum analyzer mask (relative to the top line of the display (
spectrum analyzer mask (relative to the top line of the display (
Ref
Level
)).
Syntax
To define the levels in a four-point mask:
SAN:MASK:REL:LOW:POIN:LEVEL1 <real number>
SAN:MASK:REL:LOW:POIN:LEVEL2 <real number>
SAN MASK:REL:LOW:POIN:LEVEL3 <real number>
SAN:MASK:REL:LOW:POIN:LEVEL4 <real number>
SAN:MASK:REL:LOW:POIN:LEVEL1?
Screen/field equivalent
SAN:MASK:REL:LOW:POIN:LEVEL1 controls the
Level Ofs
field on
the
Mask
menu of the SPEC ANL screen when the
EditPt
field is set to
1
, the
Edit Mask
field is set to
Lower
, and the
Mask Type
field is set to
Rel
.