Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
Chapter 1
55
GPIB Command Dictionary
AFANalyzer subsystem
:DEMPhasis:GAIN <string>
:DEMPhasis:GAIN?
These commands set/query the AF analyzer’s amplifier gain. This 
command is only effective when autoranging is turned off using the 
AFAN:RANG ‘Hold’ command.
Gain is automatically set when the AFAN:RANG ‘Auto’ command is 
used. If autoranging is enabled (AFAN:RANG ‘Auto’) and you attempt 
to change the gain setting by sending the AFAN:DEMP:GAIN 
command, the value you send will be overridden by the autoranging 
function.
Syntax
AFAN:DEMP:GAIN ‘0 dB’
 
    
‘10 dB’
 
     
‘20 dB’
 
    
‘30 dB’
AFAN:DEMP:GAIN?
Screen/field equivalent
AFAN:DEMP:GAIN sets the 
De-Emp Gain
 field on the AF ANALYZER 
screen when the 
Gain Cntl
 field is set to 
Hold
.