Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
Chapter 1
57
GPIB Command Dictionary
AFANalyzer subsystem
:DETector:SETTling <string>
:DETector:SETTling?
These commands set/query the settling time for audio measurements. 
(Use Fast when measuring signals greater than 200 Hz).
Syntax
AFAN:DET:SETT ‘Fast’
 ‘Slow’
AFAN:DET:SETT?
Screen/field equivalent
AFAN:DET SETT controls the 
Settling
 field on the AF ANALYZER 
screen.
:ELResistor <real number>
:ELResistor?
These commands set/query the external load resistance for 
measurements using the AUDIO IN HI and LO connectors. 
This command utilizes the “Real Number Setting Syntax” on page 299, 
but does not use the :STATe command.
Syntax
AFAN:ELR <real number> !valid from 1 ohm to 1 megohm
AFAN:ELR? !returns the value in ohms
Screen/field equivalent
AFAN:ELR controls the 
Ext Load R
 field on the AF ANALYZER screen.