PIXIS System Manual 
Version 2.C 
 
the problem is fixed, stop acquisition or continue to the “Focusing” topic on 
page 42; otherwise, stop data acquisition and continue to Step g.  
  If you do not hear a shutter operating and you have done Steps a-e, stop data 
acquisition and continue to Step g. 
g.  Make sure the spectrograph has an entrance slit shutter. An externally mounted 
shutter is easily confirmed. Verifying an internally mounted shutter requires 
access to the inside of the spectrograph: refer to the spectrograph manual for 
instructions. 
h.  Check to see if the PIXIS has a shutter. Loosen and remove the camera from the 
spectrograph. If you see the CCD when you look into the front of the camera, the 
camera does not have an internal shutter. Re-mount the camera to the spectrograph. 
i.  If the camera and spectrograph both have shutters, permanent damage to the 
shutter driver may occur. You can either disconnect and remove the entrance slit 
shutter or control it via a Princeton Instruments SHC-EXT Control Box. To 
disconnect and remove an entrance slit shutter: 
  External Slit Shutter: Disconnect the shutter cable and remove the shutter 
assembly from the spectrograph entrance port. 
  Internal Slit Shutter: Disconnect the shutter cable. Removing an internally 
mounted shutter requires access to the inside of the spectrograph: refer to the 
spectrograph manual for removal instructions. 
5.  If you need more help, contact Customer Support. See page 132 for contact 
information. 
Focusing 
The mounting hardware provides two degrees of freedom, focus and rotation. In this 
context, focus means to physically move the camera back and forth through the focal 
plane of the spectrograph. The approach taken is to slowly move the camera in and out of 
focus and adjust for optimum while watching a live display on the monitor, followed by 
rotating the camera and again adjusting for optimum. The following procedure, which 
describes the focusing operation with an Acton SP-2300i spectrograph, can be easily 
adapted to other spectrographs. For IsoPlane SCT-320 related focusing information, see 
IsoPlane SCT-320 Spectrograph” on page 44. 
If the PIXIS has an internal shutter, DO NOT USE the Shutter connector to drive a 
spectrograph entrance slit shutter. If both the camera and the spectrograph have 
shutters, the entrance slit shutter will have to be removed or a Princeton Instruments 
SHC-EXT Shutter Control Box may be used to control the opening and closing of the 
entrance slit shutter. See the spectrograph manual. 
Acton Series Spectrograph 
1.  Mount a light source such as a mercury pen-ray type in front of the entrance slit of 
the spectrograph. Any light source with line output can be used. Standard fluorescent 
overhead lamps have good calibration lines as well.
 If there are no "line" sources 
available, it is possible to use a broadband source such as tungsten for the alignment. 
If this is the case, use a wavelength setting of 0.0 nm for alignment purposes. 
Caution