Note: When aligning other accessories, such as fibers, lenses, optical fiber adapters, first 
align the spectrograph to the slit. Then align the accessory without disturbing the camera 
position. The procedure is identical to that used to focus the spectrograph (i.e., do the 
focus and alignment operations while watching a live image). 
10.  Tighten the spectrograph set screws to secure the spectrograph adapter. Then stop 
data acquisition. 
IsoPlane SCT-320 Spectrograph 
Because the PIXIS is mounted directly to the mounting plate on the IsoPlane, focusing 
and alignment is different from the way that focusing and alignment are performed for an 
Acton Series spectrograph. The following information assumes that you are familiar with 
the locations of the mounting plate, Micrometer Compartment, and the locking set screw. 
If not, refer to the IsoPlane manual supplied with the spectrograph. 
1.  Mount an Acton light source such as the dual HG/NeAr source in front of the 
entrance slit of the spectrograph.  
2.  With the spectrograph properly connected to the computer, turn the power on, wait 
for the spectrograph to initialize.  
3.  With the PIXIS mounted to the spectrograph and connected to the computer, turn on 
the power and wait for the detector to initialize. 
4.  Boot the WinSpec software. Select 
Define
 from the Spectrograph menu, click on 
Install/Remove Spectrograph
, highlight “Acton SCT320” in the 
Supported 
Spectrographs
 list, and click on 
Install Selected Spectrograph
.  
5.  Select 
Move
 from the Spectrograph menu, choose the grating to be move, and then 
set it to 500 nm if using a mercury lamp or to 0.0 nm if using a broadband source.  
6.  Set the slit to 10 µm at a minimum. If necessary, adjust the Exposure Time to 
maintain optimum (near full-scale) signal intensity. 
7.  Wait until the detector temperature locks at its default temperature. 
8.  Remove the cover from the Micrometer Compartment. 
9.  Using a 3/32” hex wrench, loosen the locking set screw. 
10.  Turn on 
Focus
 mode. 
11.  While continuously acquiring data, adjust the micrometer until you maximize the 
intensity level of a selected peak or peaks. 
12.  Tighten down the locking set screw. 
13.  Place the Micrometer Cover on the spectrograph. Replace and tighten all of the cover 
screws. 
14.  Next adjust the rotation. First, use a 9/64” hex wrench to loosen the four screws at the 
corners of the detector mounting plate. While watching a live display of the 
spectrum, select a peak to monitor and then rotate the detector (up to 4 degrees of 
rotation are possible). The peak will go from broad to narrow and back to broad. 
Leave the detector rotation set for the narrowest achievable peak.  
Alternatively, take an image, display the horizontal and vertical cursor bars, and 
compare the vertical bar to the line shape on the screen. Rotate the detector until the 
line shape on the screen is parallel with the vertical bar. 
15.  After completing the rotational alignment, re-tighten the four mounting plate screws. 
Then stop acquisition.