Agilent Technologies 22ET 8719ES Manuale Utente
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
1-116
Making Measurements
Using Test Sequencing to Test a Device
To create a second sequence that will perform a desired measurement function,
decrement the loop counter, and call itself until the loop counter value is equal to
zero, press:
decrement the loop counter, and call itself until the loop counter value is equal to
zero, press:
or on ET models:
This will create a displayed list as shown:
SEQUENCE LOOP 2
Start of Sequence
Trans:FWD S21 (B/R)
SCALE/DIV
AUTO SCALE
MKR Fctn
SEARCH MAX
DECR LOOP COUNTER
IF LOOP COUNTER <> 0 THEN DO
SEQUENCE 2
To run the loop sequence, press:
Generating Files in a Loop Counter Example Sequence
This example shows how to increment the names of files that are generated by a sequence
with a loop structure.
with a loop structure.
Seq
NEW SEQ/MODIFY SEQ SEQUENCE 2 SEQ2
Meas
Trans: FWD S21 (B/R)
TRANSMISSN
Scale Ref
AUTO SCALE
Marker Search
SEARCH: MAX
Seq
SPECIAL FUNCTIONS DECISION MAKING
DECR LOOP COUNTER IF LOOP COUNTER 0
SEQUENCE 2 SEQ2
Seq
DONE SEQ MODIFY
Preset
SEQUENCE 1 SEQ1
Seq
NEW SEQ/MODIFY SEQ SEQUENCE 1 SEQ 1
SPECIAL FUNCTIONS DECISION MAKING
LOOP COUNTER
7
x1
Save/Recall
SELECT DISK INTERNAL DISK
RETURN DEFINE DISK-SAVE DATA ONLY ON
Local
SET ADDRESSES PLOTTER PORT DISK
Seq
DO SEQUENCE SEQUENCE 2
DONE SEQ MODIFY
NEW SEQ/MODIFY SEQ SEQUENCE 2 SEQ 2