Agilent Technologies 22ET 8719ES Manuale Utente

Pagina di 478
2- 7
Making Mixer Measurements (Option 089 Only)
Measurement Considerations
Figure 2-5 Example of Conversion Loss versus Output Frequency with Correct 
IF Signal Path Filtering and Attenuation at All Mixer Ports 
How RF and IF Are Defined 
When you choose between 
 and 
 in the frequency offset menus, the 
analyzer determines which direction the internal source must sweep in order to achieve 
the requested IF frequency. For example, to measure the lower sideband of a mixer, where 
the RF signal is below the LO (
), the internal source must sweep backwards. See 
the examples in 
.
 RF < LO 
 RF > LO 
 RF < LO