Making Mixer Measurements (Option 089 Only)
Conversion Loss Using the Frequency Offset Mode
Performing a Power Meter (Source) Calibration Over the RF Range
1. Calibrate and zero the power meter.
2. Set the power meter’s address:
 
 (where aa is the GPIB address of the power meter) 
3. Select the appropriate power meter by pressing 
 until the correct 
model number is displayed (Agilent 436A or Agilent 438A/437).
NOTE
The Agilent E4418B and E4419B (EPM) power meters have a “437emulation” 
mode which can be used in this procedure by following these steps.
• If you are using an Agilent EPM power sensor and your network analyzer 
has firmware revision 7.72 or greater:
a. Select the remote interface command set on the power meter by 
pressing the System Inputs key and the following softkeys: Remote 
Interface
Command SetSCPI
b. Skip step 4 and continue at step 5.
The power sensor factors are automatically read by the analyzer.
• If you are using an Agilent 848X-series power sensor or your network 
analyzer does not have firmware revision 7.72 or greater:
a. Choose the 438A/437 selection on the network analyzer by pressing:
 
 until it reads 
b. Select the remote interface command set on the power meter by 
pressing the System Inputs key and the following softkeys: Remote 
Interface
Command Set437B
4. Press 
  
  
 
and 
enter the correction factors as listed on the power sensor. Press 
 
 
 (where fff is the frequency of the calibration factor in MHz) 
 
 
 (where nnn is the calibration factor number) 
 
 for each correction 
factor. When finished, press 
.
 SET ADDRESSES 
 ADDRESS: P MTR/GPIB 
aa
x1
 POWER MTR [ ] 
Local
 SET ADDRESSES 
 POWER MTR [ ] 
 POWER MTR: [438A/437] 
Cal
 PWRMTR CAL   LOSS/SENSR LISTS   CAL FACTOR   SENSOR A 
 ADD   FREQUENCY 
fff
M/
µ
 CAL FACTOR 
nnn
x1
 DONE 
 DONE