Agilent Technologies 22ET 8719ES Manuale Utente

Pagina di 478
2-32
 
Making Mixer Measurements (Option 089 Only)
Fixed IF Mixer Measurements
Sequence 2 Setup 
The following sequence makes a series of measurements until all 26 CW measurements 
are made and the loop counter value is equal to zero. This sequence includes:
• taking  data
• incrementing the source frequencies
• decrementing the loop counter
• labeling the screen
1. Press the following keys on the analyzer to create sequence 2:
NOTE
To enter the following sequence commands that require titling, an external 
keyboard may be used for convenience.
 
 
 
Taking Data
 
 
 
 
 
 
Incrementing the Source Frequencies
 
 
 
Input as title: FREQ:CW UP
 
 
 
 
 
 
 
Decrementing the Loop Counter
 
 
 
Labeling the Screen
 
 
 
Input as title: MEASUREMENT COMPLETED
 
Seq
 DONE SEQ MODIFY 
Seq
 NEW SEQ/MODIFY SEQ   SEQUENCE 2 SEQ2 
Seq
 SPECIAL FUNCTIONS   WAIT x 
.1
x1
Sweep Setup
 TRIGGER MENU   MANUAL TRG ON POINT 
Display
 MORE   TITLE   ERASE TITLE 
 DONE 
Seq
 SPECIAL FUNCTIONS   PERIPHERAL GPIB ADDR 
19
x1
 TITLE TO PERIPHERAL 
 PERIPHERAL GPIB ADDR 
21
x1
 TITLE TO PERIPHERAL 
 DECISION MAKING   DECR LOOP COUNTER   IF LOOP COUNTER<>0 
 SEQUENCE 2 SEQ2 
Display
 MORE   TITLE   ERASE TITLE 
 DONE 
Seq
 DONE SEQ MODIFY