Making Measurements
Making a Basic Measurement
Setting the Frequency Range
To set the center frequency to 134 MHz, press:
To set the span to 30 MHz, press:
NOTE
You could also press the
and
keys and enter the frequency
range limits as start frequency and stop frequency values.
Setting the Source Power
To change the power level to
−
5 dBm, press:
NOTE
You could also press
and select
one of the power ranges to keep the power setting within the defined range.
Setting the Measurement
To change the number of measurement data points to 101, press:
To select the transmission measurement, press:
or on ET models:
To view the data trace, press:
Step 3. Perform and apply the appropriate error-correction.
Refer to the
for procedures on
correcting measurement errors.
To save the instrument state and error-correction in the analyzer internal memory,
press:
press:
Step 4. Measure the device under test.
Replace any standard used for error-correction with the device under test.
To measure the insertion loss of the bandpass filter, press:
Center
134
M/
µ
Span
30
M/
µ
Start
Stop
Power
−
5
x1
POWER RANGE MAN POWER RANGES
Sweep Setup
NUMBER OF POINTS
Meas
Trans: FWD S21 (B/R)
TRANSMISSN
Scale Ref
AUTOSCALE
Save/Recall
SELECT DISK INTERNAL MEMORY RETURN SAVE STATE
Marker Search
SEARCH: MAX