Agilent Technologies 22ET 8719ES Manuale Utente

Pagina di 478
3- 15
Making Time Domain Measurements
Time Domain Low Pass Mode
Time Domain Low Pass Mode
This mode is used to simulate a traditional time domain reflectometry (TDR) 
measurement. It provides information to determine the type of discontinuity (resistive, 
capacitive, or inductive) that is present. Low pass provides the best resolution for a given 
bandwidth in the frequency domain. It may be used to give either the step or impulse 
response of the test device.
The low pass mode is less general-purpose than the bandpass mode because it places strict 
limitations on the measurement frequency range. The low pass mode requires that the 
frequency domain data points are harmonically related from dc to the stop frequency. That 
is, stop = n 
×
 start, where n = number of points.
For example, with a start frequency of 50 MHz and 101 points, the stop frequency would be 
5.05 GHz. Since the analyzer frequency range starts at 50 MHz, the dc frequency response 
is extrapolated from the lower frequency data. The requirement to pass dc is the same 
limitation that exists for traditional TDR.
Setting the Frequency Range for Time Domain Low Pass 
Before a low pass measurement is made, the measurement frequency range must meet the 
(stop = n 
×
 start) requirement previously described. The 
 softkey 
performs this function automatically: the stop frequency is set close to the entered stop 
frequency, and the start frequency is set equal to stop/n.
If the low end of the measurement frequency range is critical, it is best to calculate 
approximate values for the start and stop frequencies before pressing 
 and calibrating. This avoids distortion of the measurement 
results. To see an example, select the preset values of 201 points and a 50 MHz to 13.5 GHz 
frequency range. Now press 
 and observe the change in frequency 
values. The stop frequency changes to 13.499 GHz, and the start frequency changes to 
67.164 MHz. This would cause a distortion of measurement results for frequencies from 
50 MHz to 67.164 MHz.
NOTE
If the start and stop frequencies do not conform to the low pass requirement 
before a low pass mode (step or impulse) is selected and transform is turned 
on, the analyzer resets the start and stop frequencies. If error correction is on 
when the frequency range is changed, this turns it off. Therefore, set the 
frequency range for time domain low pass before performing a calibration.
Table 3-2  Minimum Frequency Ranges for Time Domain Low Pass 
Number of Points 
Minimum Frequency Range 
Number of Points 
Minimum Frequency Range 
3
50 MHz to 150 MHz
101
50 MHz to 5.05 GHz
11
50 MHz to 550 MHz
201
50 MHz to 10.05 GHz
21
50 MHz to 1.05 GHz
401
50 MHz to 20.05 GHz
 SET FREQ LOW PASS 
 SET FREQ LOW PASS 
 SET FREQ LOW PASS