Agilent Technologies 22ET 8719ES Manuale Utente

Pagina di 478
5- 7
Optimizing Measurement Results
Maintaining Test Port Output Power During Sweep Retrace
Maintaining Test Port Output Power During Sweep 
Retrace
During standard operation, the analyzer provides output power during its forward 
frequency sweep, but may not provide output power during its sweep retrace.  If the device 
under test (such as an amplifier with AGC circuitry) requires constant power, then you can 
set the analyzer to maintain test port output power during its sweep retrace.
To activate this feature, press 
 
 
. The analyzer 
now maintains test port output power during sweep retrace. 
NOTE
On the 8722ET/ES only, retrace power will be lost at the point when the 
sweep crosses the frequency boundary of 20 GHz.
NOTE
The frequency sweep of the network analyzer is such that when it reaches 
2.55 GHz, it momentarily jumps to 6.35 GHz before returning to 2.55 GHz 
and continuing the rest of the sweep. This discontinuity in the frequency 
sweep may be of concern to those making measurements of frequency 
sensitive amplifiers having AGC circuitry.
System
 CONFIGURE   RETRCE PWR ON