Agilent Technologies 22ET 8719ES Manuale Utente

Pagina di 478
5- 13
Optimizing Measurement Results
Increasing Sweep Speed
To View a Single Measurement Channel 
Viewing a single channel will increase the measurement speed if the analyzer’s channels 
are in alternate, or uncoupled mode.
1. Press 
   
.
2. Press 
 and 
 to alternately view the two measurement channels.
If you must view both measurement channels simultaneously (with dual channel), use 
the chop sweep mode, explained next.
3. If you want to view channel 3 (or channel 4), press 
 (or 
). This will 
always result in a dual trace display of channel 1 and channel 3 (or channel 2 and 
channel 4).
To return to a single-trace display, press 
 
 
To Activate Chop Sweep Mode 
You can use the chop sweep mode to make two measurements at the same time. For 
example, the analyzer can measure S11 and S21 simultaneously. You can activate the chop 
mode by pressing 
 or by pressing 
 
 
.
While Chop mode is the fastest way to measure devices, some components, such as filters 
with very high attenuation, may require measurement in Alternate mode. See 
.
To Use External Calibration 
Off-loading the error correction process to an external PC increases throughput on the 
network analyzer. This can be accomplished with remote-only commands. Refer to the 
programmer’s guide for information on how to use external calibration.
To Use Fast 2-Port Calibration (ES Analyzers Only)
With a 2-port calibration on, faster trace updates are possible by not measuring the 
reverse path for every forward sweep. This is controlled by the test set switch command. 
This is convenient for tuning applications because it gives a faster trace update.
When making measurements using full two-port error-correction, the following types of 
test set switching can be defined by the user:
• Hold: In this mode the analyzer does not switch between the test ports on every sweep. 
The measurement stays on the active port after an initial cycling between the ports. 
The fastest measurements can be made by using this type of test set switching. 
Pressing the 
 key, changing to a different S-parameter measurement, or any 
other action which restarts a sweep will cause the test set to switch and cycle between 
the ports.
Display
 DUAL | QUAD SETUP    DUAL CHAN on OFF 
  AUX CHAN on OFF 
Chan 1
Chan 2
Chan 3
Chan 4
Display
 DUAL | QUAD SETUP 
  AUX CHAN on OFF 
Preset
Cal
 MORE   CHOP A and B 
Meas