Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
❏ If you are modifying the frequency, enter the new value, followed by a 
or 
 key.
❏ If you are modifying the correction factor, enter the new value, followed by the 
 
key.
4. Press 
 after you have finished modifying the segment.
5. If you want to edit any other segments, press 
 and follow the previous steps, 
starting with step 2.
Deleting Frequency Segments 
1. Access the Segment Modify menu by pressing 
 
 
 
 (or 
depending on where the segment is that you want to delete).
2. Identify the segment that you want to delete by pressing 
 and using the 
 
and 
 keys to locate and position the segment next to the pointer (>), shown 
on the display. Or press 
 and enter the segment number followed by 
.
3. Press .
The analyzer deletes the segment and moves the remainder of the segments up one 
number.
4. You could also delete all the segments in a list by pressing 
 
.
5. Press 
 when you are finished modifying the segment list.
Compensating for Directional Coupler Response 
If you use a directional coupler to sample power in your measurement configuration, you 
should enter the coupled arm power loss value into the power loss table, using the 
following procedure. You can enter the loss information in a single segment, and the 
analyzer will assume that the value applies to the entire frequency range of the 
instrument. Or, you can input actual measured power loss values at several frequencies 
using up to 12 segments, enhancing power accuracy.
1. Press 
  
 .
The analyzer shows the notation 
EMPTY
, if you have not entered any segment 
information.
2. To create the first segment, press 
 
 and enter a frequency of a 
correction factor data point, followed by the appropriate key: 
 
 
.
3. Press 
 and enter the power loss that corresponds to the attenuation of the 
directional coupler (or power splitter) at the frequency that you have entered in the 
previous step. Complete the power loss entry by pressing 
 
.
G/n
M/
µ
k/m
x1
 DONE 
 SEGMENT 
Cal
 PWRMTR CAL 
 LOSS/SENSR LISTS   CAL FACTOR SENSOR A 
 CAL FACTOR SENSOR B 
 SEGMENT 
 SEGMENT 
x1
 DELETE 
 CLEAR LIST   YES 
 DONE 
Cal
 PWRMTR CAL   LOSS/SENSR LISTS   POWER LOSS 
 ADD   FREQUENCY 
G/n
M/
µ
k/m
 LOSS 
x1
 DONE