6-62
 
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
Perform the Calibration
1. Press 
 .
When ECal is first selected (or when you select module A or module B), there is a small 
initial delay so that the network analyzer can detect and download the calibration 
information from the internal memory of the ECal module.
2. Press 
 so that A is selected.
NOTE
If you are calibrating with two modules, the overlapping frequency span will 
be determined by the second module. Therefore, if you want to use the 
calibration data of Module A in the overlapping frequency span, calibrate 
using Module B first, and then calibrate using Module A.
3. Press the ECal calibration selection. 
• The calibration choices for ES models are: 
 performs a measurement calibration for reflection-only. Measures 
of one-port devices or properly terminated two-port devices, at port 1 of an 
S-parameter test set.
 performs a measurement calibration for reflection-only. Measures 
of one-port devices or properly terminated two-port devices, at port 2 of an 
S-parameter test set.
 performs a complete calibration for measurement of all four 
S-parameters of a two-port device. This is the most accurate calibration for 
measurements of two-port devices.
 performs an S11 and S21 enhanced response calibration 
(forward direction). Enhanced response generates a 1-port cal for S11 and an 
improved calibration over the response cal for S21.
 performs an S22 and S12 enhanced response calibration 
(reverse direction). Enhanced response generates a 1-port cal for S22 and an 
improved calibration over the response cal for S12.
• The calibration choices for ET models are: 
 performs a reflection 1-port calibration.
 performs a transmission (same as S21) and reflection 
(same as S11) enhanced response calibration (forward direction). Enhanced 
response generates a 1-port cal for reflection and an improved calibration over the 
response cal for transmission. 
Once the ECal calibration selection is made, the network analyzer performs the 
calibration.
4. If you selected the manual thru calibration option, when the prompt is displayed, 
remove the ECal module from the setup and connect the manual thru (the two test 
ports connected together) as shown in 
.
Cal
 ECal MENU 
 MODULE A b 
 S11 1-PORT 
 S22 1-PORT 
 FULL 2-PORT 
 S11/S21 ENH. RESP. 
 S22/S12 ENH. RESP. 
 REFLECTION 1-PORT 
 TRANS/REFL ENH. RESP.