Operating Concepts
Output Power
Output Power
Understanding the Power Ranges 
The built-in synthesized source contains a programmable step attenuator that allows you 
to directly and accurately set power levels in twelve different power ranges. Each range 
has a total span of 20 dB (15 dB, 8722ET/ES). The twelve ranges cover the instrument’s 
full operating range. In addition, some amount of overrange and underrange is permitted 
beyond the stated limits. Performance outside of the range limits varies from one analyzer 
to another and is not specified. A power range can be selected either manually or 
automatically.
Automatic mode 
If you select 
, you can enter any power level within the total 
operating range of the instrument and the source attenuator will automatically switch to 
the corresponding range. 
Each range overlaps its adjacent ranges by 15 dB, therefore, certain power levels are 
designated to cause the attenuator to switch to the next range so that optimum (leveled) 
performance is maintained. These transition points exist at 
10 dB from the top of a range 
and at +5 dB from the bottom of a range. This leaves 10 dB of operating range. By turning 
the RPG knob with 
 being the active function, you can hear the attenuator 
switch as these transitions occur. See the specifications and characteristics chapter in the 
reference guide for the analyzer’s power limits.
Manual mode 
If you select 
, you must first enter the power ranges menu and 
manually select the power range that corresponds to the power level you want to use. This 
is accomplished by pressing the 
 softkey and then selecting one of the 
12 available ranges. The active power will always be reset to be within this range. In this 
mode, you will be able to select power levels above or below the range limits. The factory 
limits on setting power level are +6 dB above the maximum range power and 
12 dB below 
the minimum range power. Typical performance for power overrange is 2 dB above the 
maximum (more at lower frequencies and less at higher frequencies) and typical under 
range is 12 dB. An under-range condition may cause a phase lock lost message to be 
displayed.
When a calibration is active, the power range selection is switched from auto to manual 
mode, and 
PRm
 appears on the display. This feature is necessary to maintain accuracy once 
a measurement calibration is activated.
PWR RANGE AUTO 
PORT POWER 
PWR RANGE MAN 
POWER RANGES