Operating Concepts
Calibration Routines
Calibration Routines
There are twelve different error terms for a two-port measurement that can be corrected 
by accuracy enhancement in the analyzer. These are directivity, source match, load match, 
isolation, reflection tracking, and transmission tracking, each in both the forward and 
reverse direction. The analyzer has several different measurement calibration routines to 
characterize one or more of the systematic error terms and remove their effects from the 
measured data.
The calibrate menu allows you to perform the measurement calibration routines. These 
procedures range from a simple frequency response calibration to a full two-port 
calibration that effectively removes all twelve error terms.
Response Calibration 
The response calibration, activated by pressing the 
 softkey within the 
calibrate menu, provides a normalization of the test setup for reflection or transmission 
measurements. This calibration procedure may be adequate for measurement of well 
matched devices. This is the simplest error-correction to perform, and should be used when 
extreme measurement accuracy is not required.
Response and Isolation Calibration 
The response and isolation calibration, activated by pressing the 
 
softkey within the calibrate menu, provides a normalization for frequency response and 
crosstalk errors in transmission measurements, or frequency response and directivity 
errors in reflection measurements. This procedure may be adequate for measurement of 
well matched high-loss devices.
Enhanced Response Calibration 
The enhanced response calibration, activated by pressing the 
 
softkey within the calibrate menu, provides a one-port calibration to correct for directivity, 
source match, and frequency response for reflection measurements, and corrects for source 
match as well as frequency response for transmission measurements. 
For ES models, enhanced response calibration improves accuracy in transmission 
measurements compared to a response calibration or a response and isolation calibration, 
but it is not as accurate as a full 2-port calibration. For ET models, this calibration 
provides the best accuracy for measurements with a T/R test set.
RESPONSE 
RESPONSE & ISOL’N 
 ENHANCED RESPONSE