Agilent E1441A SCPI Command Reference
Chapter 3
Parameters
FREQuency:STARt?
[SOURce:]FREQuency:STARt? [MIN|MAX]
queries the start frequency. Returns a
value in hertz.
FREQuency:STOP
[SOURce:]FREQuency:STOP <
frequency
>|MIN|MAX
sets the stop frequency.
Select from 10 mHz to 15 MHz (100 kHz for triangle and ramp). The sweep is phase
continuous over the full frequency range. The default is 1 kHz. MIN = 10 mHz.
continuous over the full frequency range. The default is 1 kHz. MIN = 10 mHz.
MAX
= 15 MHz. [ Stored in volatile memory ]
Parameters
FREQuency:STOP?
[SOURce:]FREQuency:STOP? [MIN|MAX]
queries the stop frequency. Returns a
value in hertz.
Frequency-Shift Keying (FSK) Commands
Use the APPLy command or the equivalent FUNC:SHAP, FREQ, VOLT, and
VOLT:OFFS commands to configure the carrier waveform. Set the carrier frequency
between 100
VOLT:OFFS commands to configure the carrier waveform. Set the carrier frequency
between 100
µHz and 15 MHz (100 kHz maximum for triangle and ramp). The
default is 1 kHz.
Parameter Name
Parameter Type
Range of Values
Default Units
frequency
numeric
10E-03 to 15E+06
Hz
Parameter Name
Parameter Type
Range of Values
Default Units
frequency
numeric
10E-03 to 15E+06
Hz