Agilent E1441A SCPI Command Reference  
Chapter 3
When the external source is selected, the output frequency is determined by the 
signal level on the source selected by the TRIG:SOUR command. When a 
“low” 
TTL
 level is present, the carrier frequency is output. When a “high” 
TTL
 
level is present, the hop frequency is output. 
The maximum external 
FSK
 rate is 1 MHz.
FSKey:SOURce?
[SOURce:]FSKey:SOURce? 
queries the internal 
FSK
 source. Returns “
INT
” or 
EXT
”.
FSKey:STATe
[SOURce:]FSKey:STATe OFF|ON 
disables or enables 
FSK
 modulation. To ensure 
proper operation, you should enable 
FSK
 after you have set up the other modulation 
parameters. Only one modulation mode can be enabled at a time. When you enable 
FSK
, the previous modulation mode is turned off. 
FSKey:STATe?
[SOURce:]FSKey:STATe? 
queries the state of the 
FSK
 mode. Returns “0” (
OFF
) or 
“1” (
ON
). 
Selecting an Arbitrary Waveform
FUNCtion:USER
[SOURce:]FUNCtion:USER <
arb name
>|VOLATILE 
selects one of the five built-in 
arbitrary waveforms, one of four user-defined waveforms, or the waveform currently 
downloaded to 
VOLATILE
 memory. 
Parameters
The arb name may contain up to 8 characters. The first character must be a 
letter (A-Z), but the remaining characters can be numbers (0-9) or the 
underscore character (“ _ ”). Blank spaces are not allowed. If you specify a 
name with more than 8 characters, a +783, “Arb waveform name too long” 
error is generated. 
The function generator does not distinguish between upper- and lower-case 
letters for the arbitrary waveform name. Therefore, ARB_1 and arb_1 are the 
same name. All characters are converted to upper case.