Service Procedures  
Appendix D
AM Modulation Depth Verification
This test verifies the modulation depth specification.
Select each function generator output in the table below. Use a DMM to 
measure the function generator ACrms output voltage. Compare the 
measured results to the test limits shown in the table. This is a HIGH Z output 
termination test.
Optional Performance Verification Tests
These tests are not intended to be performed with every calibration. They are 
provided as an aid for verifying additional instrument specifications.
Square Wave Duty
Cycle Verification
This test verifies the duty cycle specification of the squarewave output.
Select each function generator output in the table below. Use an integrating 
DMM to measure the VDC output of the function generator. Compare the 
measured results to the test limits shown in the table. This is a HIGH Z output 
termination test.
NOTE:  You make a DC voltage measurement on this test.
Note
The DMM used for this test must be an integrating multimeter. If the first 
step does not measure 0 VDC, use an oscilloscope for this test.
Do not use an auto-ranging function on the DMM for this test. Fix the 
DMM measurement range at 10 VDC. 
Table D-8. AM Modulation Verification
Agilent E1441A
Measurement
AM Modulation
+
Function
Out
Term
a
Ampl
Freq
Shape
Freq
Depth
Nominal
Error
Q
Sine wave
HIGH Z
1.0 Vrms
1.00 kHz
Sinewave
100 Hz
0%
0.50 Vrms
±
 0.005 Vrms
Sine wave
HIGH Z
1.0 Vrms
1.00 kHz
Sinewave
100 Hz
100%
0.61 Vrms
±
 0.0061 Vrms
a.The E1441A has a fixed output impedance of 50 
Ω
 on the "Output" terminal.  Use the OUTPut:LOAD 
50 | INFinity | MIN | MAX command to set the output termination. HIGH Z assumes no load on output. 
50 
Ω
 assumes a 50 
Ω
 
±
 0.1 
Ω
 load on output.
Table D-9. Square Wave Duty Cycle Verification
Function
 Out
Term
a
Ampl
Freq
Duty 
Cycle
Nominal
Error 
O
Square wave
HIGH Z
1.0 Vrms
300.00 Hz
50%
0.00 VDC
±
 0.020 VDC
O
Square wave
HIGH Z
1.0 Vrms
300.00 Hz
25%
- 0.50 VDC
±
 0.020 VDC
O
Square wave
HIGH Z
1.0 Vrms
300.00 Hz
75%
+ 0.50 VDC
±
 0.020 VDC
a.The E1441A has a fixed output impedance of 50 
Ω
 on the "Output" terminal.  Use the 
OUTPut:LOAD 50 | INFinity | MIN | MAX command to set the output termination. 
HIGH Z assumes no load on output. 50 
Ω
 assumes a 50 
Ω
 
±
 0.1 
Ω