Amplitude Flatness Verification
This test verifies the output amplitude flatness specification at selected 
frequencies. If you use a Temperature Voltage Converter (TVC) (the 
recommended method) 
or a wide band ACrms voltmeter (with a 50 
Ω  feed through load), perform 
this procedure as described. If you are using a measurement device that 
requires a transfer measurement (for example, a power meter), make the 
transfer in the reference measurement at 100 kHz. 
Set the function generator to the first output indicated in the table below and 
make a reference measurement. Set each function generator output shown in 
the table below.  Measure the output of the function generator. Compare the 
amplitude level measured to the reference measurement +/- the error shown 
in the table. This test is a 50 
Ω output termination test.
Table D-7. Amplitude Flatness Verification
Agilent E1441A
Measurement
Function
 Out
Term
a
Ampl
Freq
Nominal
Error 
Q
Sine wave
50
 Ω
3.0 Vrms
1.00 kHz
<reference>
Sine wave
50 
Ω
3.0 Vrms
100.00 kHz
<reference>
 ±
 0.03 Vrms
Sine wave
50
 Ω
3.0 Vrms
500.00 kHz
<reference>
±
 0.045 Vrms
Q
Sine wave
50
 Ω
3.0 Vrms
1.00 MHz
<reference>
±
 0.045 Vrms
Sine wave
50
 Ω
3.0 Vrms
3.00 MHz
<reference>
±
 0.06 Vrms
Sine wave
50
 Ω
3.0 Vrms
5.00 MHz
<reference>
±
 0.06 Vrms
Sine wave
50
 Ω
3.0 Vrms
7.00 MHz
<reference>
±
 0.06 Vrms
Sine wave
50
 Ω
3.0 Vrms
9.00 MHz
<reference>
±
 0.06 Vrms
Sine wave
50
 Ω
3.0 Vrms
11.00 MHz
<reference>
±
 0.06 Vrms
Sine wave
50
 Ω
3.0 Vrms
13.00 MHz
<reference>
±
 0.06 Vrms
Q
Sine wave
50
 Ω
3.0 Vrms
15.00 MHz
<reference>
±
 0.06 Vrms
a.The E1441A has a fixed output impedance of 50 
Ω
 on the "Output" terminal.  
Use the OUTPut:LOAD 50 | INFinity | MIN | MAX command to set the output 
termination. HIGH Z assumes no load on output. 50 
Ω
 assumes a 50 
Ω
 
±
 0.1 
Ω
 
load on output.