Service Procedures  
Appendix D
AC Amplitude Adjustment (50 Ohms)
1. The function generator stores sixteen calibration constants related to 
50 
Ω output. The constants are calculated from the adjustment value 
entered. The calibration constants are stored following completion of 
SETUP 49 and the calibration procedure may be aborted after that 
point. No calibration constants are stored if the procedures are 
aborted at any other setup. 
2. Use the DMM to measure the resistance of a 50 
Ω feedthrough load. 
Record the measurement for step 3. You can measure the load and 
cable resistance (recommended procedure) or just the load as shown 
below.
3. Enter the following SETUP 33 and measure the value of the 50 
Ω 
feedthrough load (and cable). Send the CAL:VALue <measured 
Ω> 
to the function generator. This number will be used to calculate the 50 
Ω output amplitude calibration constants.
Note
Once the value of the 50 
Ω load and cable are entered, use the SAME load 
and cable for all 50 
Ω tests. 
4. Use the DMM to measure the function generator ACrms output 
voltage for SETUP 34 in the table on the next page. These 
adjustments use the 50 
Ω load and cable measured in step 2 and  
connected as shown in Figure D-3.
Figure D-2. 50 Ohm Feedthrough/Cable Resistance Measurement
Table D-15. 50 Ohm Adjustment Setup
Nominal Input
SETUP
LOAD Z
33
 50 
Ω
Enter measured value of load.