184
Service Procedures  
Appendix D
DC Output Adjustment
The function generator stores nine calibration constants related to DC volts 
output. The constants are calculated from the adjustment value entered. The 
calibration constants are stored following completion of SETUP 59. No 
calibration constants are stored if the procedures are aborted at any other 
setup.
 
1. Use a DMM to measure the function generator DCV output voltage 
for SETUP 50 in the following table. These adjustments use a HIGH 
Z output termination.
2. Send the CAL:VALue <measured Vdc> to the function generator. 
3. Send the CAL? command to the function generator to initiate the 
calibration.
4. Repeat steps 1, 2 and 3 for SETUP 51 through 59.
5. Perform the “DC Function Offset Verification” on page 168.
Table D-17. DC Output Adjustment Setups
Nominal Output
SETUP
DC Volts
Adjustment for: 
50
- 8.0 VDC
Negative offset gain
51
8.0 VDC
Positive offset gain
52
0.0 VDC
AM offset 
53
0.0 VDC
2 dB Pre-attenuator offset.
54
0.0 VDC
4 dB Pre-attenuator offset.
55
0.0 VDC
6 dB Pre-attenuator offset.
56
0.0 VDC
8 dB Pre-attenuator offset.
57
0.0 VDC
10 dB Pre-attenuator offset.
58
0.0 VDC
12 dB Pre-attenuator offset.
59
0.0 VDC
14 dB Pre-attenuator offset.