Agilent E1441A Application Information  
Chapter 2
Modulation depth:
0% to 120%. The default is 100%. 
Use the following command to set the modulation depth:
AM:DEPTh <depth in percent>|MINimum|MAXimum
Amplitude Modulating
Source
The function generator will accept an internal modulating signal, an external 
modulating signal, or both. 
Modulating source:
Internal-External (both) or External only. The default 
is Both (internal-external).
The External modulating source is always enabled. 
When both sources are enabled (internal-external), the function generator 
adds the internal and external modulating signals (the carrier waveform is 
actually modulated with two waveforms).
When the internal source is disabled (external only), the carrier waveform is  
expecting a modulating signal on the AM Modulation terminal.
You apply the external modulating waveform to the AM Modulation 
terminal. The modulation depth is controlled by the signal level present 
(5 volts 
peak
 corresponds to 100% modulation).
Use the following command to set the modulating source:
AM:SOURce BOTH|EXTernal
Frequency
Modulation (FM)
A modulated waveform consists of a carrier waveform and a modulating 
waveform
. In 
FM
, the frequency of the carrier is varied by the amplitude of 
the modulating waveform. The function generator will accept only an 
internal 
FM
 modulating signal (no external source is available).  Topics 
covered on frequency modulation are:
FM Carrier Waveform Shape
Figure 2-2. AM Modulation Input Signal