Agilent E1441A Application Information  
55
Chapter 2
System-Related Operations
This section gives information on topics such as error conditions and 
self-test. This information is not directly related to waveform generation but 
is an important part of operating the function generator.  Topics covered on 
system-related operations are:
Error Conditions
Self-Test
Memory State
Firmware Revision Query
SCPI Language Version Query
Error Conditions
When the faceplate 
Errors
 LED turns on, one or more command syntax or 
hardware errors have been detected. A record of up to 20 errors can be stored 
in the function generator's error queueSee Appendix B, “Error Messages,” 
for a complete listing of the errors.
Errors are retrieved in first-in-first-out (
FIFO
) order. The first error returned 
is the first error that was stored. When you have read all errors from the 
queue, the 
Errors
 LED turns off.
If more than 20 errors have occurred, the last error stored in the queue (the 
most recent error) is replaced with -350, “Too many errors”. No additional 
errors are stored until you remove errors from the queue. If no errors have 
occurred when you read the error queue, the function generator responds 
with +0, “No error”.
The error queue is cleared when power has been off or after a *CLS (clear 
status) command has been executed. The *RST (reset) command does not 
clear the error queue. 
Use the following command to check for system errors:
SYSTem:ERRor?
Reads one error from the error 
queue
Errors have the following format (the error string may contain up to 80 
characters
):
-113,"Undefined header"
Self-Test
power-on self-test occurs automatically when you turn on the function 
generator. This limited test assures you that the function generator is 
operational. 
complete self-test runs a series of tests and takes approximately 
10 seconds to execute. If all tests pass, you can have a high confidence that 
the function generator is fully operational.
If the self-test fails, the 
Errors
 LED turns on  the function generator should 
be returned to Agilent Technologies for service.