Agilent E1441A Application Information  
57
Chapter 2
Power-On and
Reset State
Note
The power-on state will state will be different if you have enabled the 
auto-recall mode. 
See “MEMory:STATe:RECall:AUTO OFF | ON | 0 | 1” 
on page 85.
Feature
Power-On and Reset State
Output Configuration
Function
Sine wave
Frequency
1 kHz
Amplitude (into 50 ohms)
100 mV peak-to-peak
Offset
0.00 Vdc
Output Units
Volts peak-to-peak
Output Termination
50 ohms
Modulation
AM Carrier Waveform
1 kHz Sine wave
AM Modulating Waveform
100 Hz Sine wave
AM Depth
100%
FM Carrier Waveform
1 kHz Sine wave
FM Modulating Waveform
10 Hz Sine wave
FM Peak Frequency Deviation
100 Hz
Burst Carrier Frequency
1 kHz Sine wave
Burst Count
1 cycle
Burst Rate
100 Hz
Burst Starting Phase
0 degrees
FSK Carrier Frequency
1 kHz Sine wave
FSK"Hop" Frequency
100 Hz Sine wave
FSK Rate
10 Hz
Modulation State
Off
Sweep Start / Stop Frequency
100 Hz / 1 kHz
Sweep Time
1 second
Sweep Mode
Linear
Triggering Operations
Trigger Source
Internal
Calibration
Calibration State
Doesn’t change with *RST or 
power-down. See page 74