Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
5- 18
Making Measurements
Math Scale and Offset
Math Scale and Offset
Any math function may be manually scaled by pressing the 
Settings softkey, and 
then adjusting the Scale or Offset value.
1 Press the 
Settings
 softkey in the Math menu to set your own scale factors 
(units/division) or offset (units) for the selected math function.
Units for each input channel can be set to Volts or Amps using the channel 
Probe 
Units softkey. Scale and offset units are:
A scale unit of 
U (undefined) will be displayed for math function 1-2 and for 
d/dt, and 
dt when 1-2 or 1+2 is the selected source if channel 1 and channel 2 
are set to dissimilar units in the channel 
Probe Units softkey.
2 Press the 
Scale
 or 
Offset
 softkey, then turn the Entry knob to rescale or 
change the offset value for your math function.
Math Scale and Offset are Set Automatically
Any time the currently displayed math function definition is changed, the function is 
automatically scaled for optimum vertical scale and offset.  If you manually set scale 
and offset for a function, select a new function, then select the original function, the 
original function will be automatically rescaled.
Math function
Units
FFT
dB
(decibels)
1*2
V
2
, A
2
, or W (Volt-Amp)
1-2
V or A
d/dt
V/s or A/s (V/second or A/second)
dt
Vs or As (V-seconds or A-seconds)
*
When the FFT source is channel 1 or channel 2 (or channel 3 or 4 on the 54624A), FFT units 
will be displayed in dBV when channel units is set to Volts and channel impedance is set to 
1 M
Ω
. FFT units will be displayed in dBm when channel units is set to Volts and channel 
impedance is set to 50
Ω
 (54640-series only).  FFT units will be displayed as dB for all other FFT 
sources or when a source channel’s units has been set to Amps.