Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
5- 20
Making Measurements
Subtract
Subtract
When you select 
1 – 2, channel 2 voltage values are subtracted from channel 1 
voltage values point by point, and the result is displayed.
You can use 1 – 2 to make a differential measurement or to compare two 
waveforms.  You may need to use a true differential probe if your waveforms 
have DC offsets larger than the dynamic range of the oscilloscope's input 
channel.
To perform the addition of channel 1 and channel 2, select 
Invert in the Channel 
2 menu and perform the 1 – 2 math function.
• Press the 
Math
 key, press the 
1 – 2
 softkey, then press the 
Settings
 softkey 
if you want to change the scaling or offset for the subtract function.
Scale allows you to set your own vertical scale factors for subtract, expressed 
as V/div (Volts/division) or A/div (Amps/division).  Units are set in the channel 
Probe menu.  Press the Scale softkey, then turn the Entry knob to rescale 1 – 2.
Offset allows you to set your own offset for the 1 – 2 math function. The offset 
value is in Volts or Amps and is represented by the center graticule of the display.  
Press the 
Offset softkey, then turn the Entry knob to change the offset for 1 – 2.
A scale unit of 
U (undefined) will be displayed for scale and offset if channel 1 
and channel 2 are set to dissimilar units in the channel 
Probe Units softkey.
The figure below shows an example of subtract.
Subtract 
Channel 1
Channel 2
1 – 2 waveform 
1 – 2 
Scale
Return to 
previous menu
1 – 2 
Offset