Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
5-43
Making Measurements
Making Delay and Phase Measurements
Making Delay and Phase Measurements
Delay
Delay measures the time difference from the selected edge on source 1 and the 
selected edge on source 2 closest to the trigger reference point at the 50% points 
on the waveforms. Negative delay values indicate that the selected edge of 
source 1 occurred after the selected edge of source 2. 
• Press the 
Settings
 softkey to select the source channels and slope for the 
delay measurement.
The default Delay settings measure from the rising edge of channel 1 to the 
rising edge of channel 2.
The example below shows a delay measurement between the rising edge of 
channel 1 and the falling edge of channel 2.
Delay Measurement
Digital channel measurements
Automatic measurements Phase, and Delay are not valid for digital channels on the 
mixed-signal oscilloscope or for the math FFT function.  The two sources defined in 
the phase and delay measurement must be turned.
Source 1
Source 2
Delay