Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
5-45
Making Measurements
Making voltage measurements automatically
Making voltage measurements automatically
Measurement units for each input channel can be set to Volts or Amps using 
the channel 
Probe Units softkey.  A scale unit of U (undefined) will be displayed 
for math function 1-2 and for d/dt, and 
dt when 1-2 or 1+2 is the selected source 
if channel 1 and channel 2 are set to dissimilar units in the channel 
Probe Units 
softkey.
The following figure shows the voltage measurement points.
Math Measurements and Units
Only Peak-Peak, Maximum, Minimum, Average, X at Min, and X at Max automatic 
measurements may be made on a math FFT function.  See “Making time 
measurements automatically” for the FFT  X at Max and X at Min measurement.  Use 
the cursors to make other measurements on FFT.  All voltage measurements may be 
made on other math functions.  The resultant units are:
FFT:
1 * 2:
1 – 2:
d/dt:
dt:
dB* (decibels)
V
2
, A
2
 or W (Volt-Amp)
V (Volts) or A (Amps)
V/s  or A/s (V/second A/second)
Vs or As (V-seconds or A-seconds)
*
When the FFT source is channel 1 or channel 2 (or channel 3 or 4 on the 54624A), FFT units 
will be displayed in dBV when channel units is set to Volts and channel impedance is set to 
1 M
Ω
. FFT units will be displayed in dBm when channel units is set to Volts and channel 
impedance is set to 50
Ω
 (54640-series only).  FFT units will be displayed as dB for all other FFT 
sources or when a source channel’s units has been set to Amps.
Maximum
Minimum
Peak-Peak
Amplitude
Top
Base