Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
2-13
Front-Panel Overview
To use digital channels to view a signal
To use digital channels to view a signal
• To configure the oscilloscope quickly, press the 
Autoscale
 key to display 
the connected signals.
• To undo  the effects of Autoscale, press the 
Undo Autoscale
 softkey before 
pressing any other key.
This is useful if you have unintentionally pressed the 
Autoscale key or do not 
like the settings Autoscale has selected and want to return to your previous 
settings.
• To set the instrument to the factory-default configuration, press the 
Save/Recall
 key, then press the 
Default Setup
 softkey.
Example 
Install probe clips on channels 0 and 1 on the digital probe cable. Connect the 
probes for digital channels 0 and 1 to the Probe Comp output on the front panel 
of the instrument. Be sure to connect the ground lead. Set the instrument to 
the factory default configuration by pressing the 
Save/Recall key, then the 
Default Setup softkey. Then press the Autoscale key. You should see a display 
similar to the following.
Autoscale with digital channels (54621D, 54622D, 54641D, and 54642D)