Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
2-15
Front-Panel Overview
To apply the default factory configuration
To apply the default factory configuration
• To set the instrument to the factory-default configuration, press the 
Save/Recall
 key, then press the 
Default Setup
 softkey.
The default configuration returns the oscilloscope to its default settings. This 
places the oscilloscope in a known operating condition. The major default 
settings are:
Horizontal  main mode, 100 us/div scale, 0 s delay, center time reference
Vertical (Analog)  Channel 1 on, 5 V/div scale, dc coupling, 0 V position, 1 M
Ω
 
impedance, probe factor to 1.0 if an AutoProbe probe is not connected to the 
channel
Trigger  Edge trigger, Auto sweep mode, 0 V level, channel 1 source, dc 
coupling, rising edge slope, 60 ns holdoff time
Display  Vectors on, 20% grid intensity, infinite persistence off
Other   Acquire mode normal, Run/Stop to Run, cursors and measurements off
Labels  all custom labels in the Label Library are erased