2-19
Front-Panel Overview
To operate the time base controls
To operate the time base controls
The following exercise guides you through the time base keys, knobs, and status 
line.
• Turn the Horizontal sweep speed (time/division) knob and notice the 
change it makes to the status line.
The sweep speed knob changes the sweep speed from 5 ns/div to 50 s/div for 
the 54620-series and 1 ns/div to 50 s/div for the 54640-series in a 1-2-5 step 
sequence, and the value of the sweep speed is displayed in the status line at the 
top of the display.
• Press the 
Main/Delayed
, then press the 
Vernier
 softkey.
The 
Vernier softkey allows you to change the sweep speed in smaller increments 
with the time/div knob. These smaller increments are calibrated, which results 
in accurate measurements, even with the vernier turned on.
• Turn the delay time knob (
) and notice that its value is displayed in 
the status line.
The delay knob moves the main sweep horizontally, and it pauses at 0.00 s, 
mimicking a mechanical detent.  At the top of the graticule is a solid triangle 
(
) symbol and an open triangle (
) symbol.  The 
 symbol indicates the trigger 
point and it moves with the Delay time knob.  The 
 symbol indicates the time 
reference point and is also where the zoom-in/zoom-out is referenced.  If the 
Time Ref softkey is set to Left, the 
 is located one graticule in from the left side 
of the display.  If the 
Time Ref softkey is set to Center, the 
 is located at the 
center of the display.  If the 
Time Ref softkey is set to Right, the 
 is located one 
graticule in from the right side of the display.  The delay number tells you how 
far the time reference point 
 is located from the trigger point 
.
All events displayed left of the trigger point 
 happened before the trigger 
occurred, and these events are called pre-trigger information.  You will find this 
feature very useful because you can now see the events that led up to the trigger 
point.  Everything to the right of the trigger point 
 is called post-trigger 
information.  The amount of delay range (pre-trigger and post-trigger 
information) available depends on the sweep speed selected.