2- 20
Front-Panel Overview
To start and stop an acquisition
To start and stop an acquisition
• When the 
Run/Stop
 key is illuminated in green, the oscilloscope is in 
continuous running mode.
You are viewing multiple acquisitions of the same signal similar to the way an 
analog oscilloscope displays waveforms.
• When the 
Run/Stop
 key is illuminated in red, the oscilloscope is stopped.
“Stop” is displayed in the trigger mode position in status line at the top of the 
display.  You may now pan and zoom the stored waveform by turning the 
Horizontal and Vertical knobs.
The stopped display may contain several triggers worth of information, but only 
the last trigger acquisition is available for pan and zoom.  To ensure the display 
does not change, use the 
Single key to be sure you have acquired only one 
trigger.
• The 
Run/Stop
 key may flash between a user requesting a 
Stop
 and the 
completion of the current acquisition.
To make a single acquisition
The Single run control key lets you view single-shot events without subsequent 
waveform data overwriting the display. Use Single when you want maximum 
memory depth for pan and zoom and you want the maximum sample rate.
1 First set trigger 
Mode/Coupling
 
Mode
 softkey to 
Normal
.
This keeps the oscilloscope from autotriggering immediately.
2 If you are using the analog channels to capture the event, turn the 
Trigger Level knob to the trigger threshold where you think the trigger 
should work.
3 To begin a single acquisition, press the 
Single
 key.
When you press Single, the display is cleared, the trigger circuitry is armed, the 
Single key is illuminated, and the oscilloscope will wait until a trigger condition 
occurs before it displays a waveform.
When the oscilloscope triggers, the single acquisition is displayed and the 
oscilloscope is stopped (
Run/Stop key is illuminated in red). Press Single again 
to acquire another waveform.