Triggering the Oscilloscope
To set holdoff
To set holdoff 
1 Press the 
Mode/Coupling
 key.
2 Turn the Entry knob 
 to increase or decrease the trigger holdoff 
time shown in the 
Holdoff
 softkey.
Holdoff sets the amount of time that the oscilloscope waits before re-arming the 
trigger circuitry. Use Holdoff to stabilize the display of complex waveforms.
To get a stable trigger on the pulse burst shown below, set the holdoff time to 
be >200 ns but <600 ns.
By setting the Holdoff, you can synchronize triggers. The oscilloscope will  
trigger on one edge of the waveform, and ignore further edges until the holdoff 
time expires. The oscilloscope will then re-arm the trigger circuit to search for 
the next edge trigger. This allows the oscilloscope to trigger on a repeating 
pattern in a waveform.