Triggering the Oscilloscope
External Trigger Input
External Trigger Input
External Trigger can be used as a source in several of the trigger types. On 
4-channel and mixed-signal oscilloscopes, the external trigger BNC input is on 
the rear panel and is labeled 
Ext Trig.  On 2-channel oscilloscopes, the external 
trigger BNC input is on the front panel and is labeled 
Ext Trigger.
Maximum input voltage for the external trigger input:
CAT I 300 Vrms, 400 Vpk
CAT II 100 Vrms, 400 Vpk
with 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
C A U T I O N    
Do not exceed 5 Vrms in 50
Ω
 mode on the 54640-series models. Input 
protection is enabled in 50
Ω
 mode and the 50
Ω
 load will disconnect if greater 
than 5 Vrms is detected. However the input could still be damaged, depending 
on the time constant of the signal.
C A U T I O N    
The 50
Ω
 input protection mode on the 54640-series models only functions 
when the oscilloscope is powered on. 
External Trigger Probe Settings
You can set probe attenuation, units, and input impedance for the external 
trigger input.
1 Press the 
Mode/Coupling
 key in the Trigger section of the front panel.