Triggering the Oscilloscope
To use Edge triggering
To use Edge triggering
The Edge trigger type identifies a trigger by looking for a specified slope and 
voltage level on a waveform. You can define the trigger source and rising or 
falling edge in this menu.   The trigger type, source, and level is displayed in the 
upper-right corner of the display.
1 Press the 
Edge
 key in the Trigger section of the front panel to display 
the Edge trigger menu.
2 Press the slope softkey (
) and select either rising edge or falling 
edge. This determines whether the trigger will occur on the rising edge 
or the falling edge of the input signal.
Slope  Selects rising edge (
) or falling edge (
) for the selected trigger 
source. The selected slope is displayed in the upper-right corner of the display.
3 Select 1, 2, Ext, or Line as the trigger source.
The trigger source can also be set to channel 3 and 4 on the 4-channel 
oscilloscope, or can be set to digital channels D15 through D0 on the 
mixed-signal oscilloscope.  You can choose a channel that is turned off as the 
source for the edge trigger.  The selected trigger source is displayed in the 
upper-right corner of the display next to the slope symbol:
1 through 4 = analog channels
D15 through D0 = digital channels
E = External trigger
L = Line trigger
Slope
Analog channel 
source
Digital channel 
source
External or Line 
source
Trigger level 
or threshold
Trigger 
source
Slope