Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
3- 14
Triggering the Oscilloscope
To use Pulse Width triggering
To use Pulse Width triggering 
Pulse Width (glitch) triggering sets the oscilloscope to trigger on a positive pulse 
or a negative pulse of a specified width. If you want to trigger on a specific 
timeout value, use 
Duration trigger in the trigger More menu.
1 Press the 
Pulse Width
 key in the Trigger section of the front panel to 
display the Pulse Width trigger menu.
2 Press the 
Source
 softkey (or rotate the Entry knob on mixed-signal 
oscilloscopes) to select a channel source for the trigger.
The channel you select is shown in the upper-right corner of the display next 
to the polarity symbol.
The source can be any analog or digital channel available on your oscilloscope.  
External trigger may also be specified as a source when using the 2-channel and 
4-channel oscilloscopes. 
Adjust the trigger level for the selected analog channel source by turning the 
Trigger 
Level knob. The position of the trigger level for the analog channel is 
indicated by the trigger level icon 
 (if the analog channel is on) at the far left 
side of the display when DC coupling is selected.  Set the threshold for the 
selected digital channel group by pressing the 
D7 Thru D0 or D15 Thru D8  key 
then press the
 Threshold softkey.  The value of the trigger level or digital 
threshold is displayed in the upper-right corner of the display.
Source
Pulse 
polarity
Qualifier
Qualifier time set
Trigger level 
or threshold
Trigger 
source
Polarity
T