Siemens Software version syngo CT 2005C ユーザーズマニュアル

ページ / 321
Abdomen
115
 
Hints
• You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for 
repeat“, e.g., when both non-contrast and contrast 
studies are required.
• Delayed scans may be required for the kidneys & 
bladder.
AbdRoutine
kV
130
Effective mAs
60
Rotation time
0.8/1.0 sec
Slice collimation
4.0 mm
Slice width
5.0 mm
Feed/Rotation
14.4 mm
Pitch Factor
1.8
Increment
5.0 mm
Kernel
B41s
CTDI
Vol
6.47 mGy
Effective dose
Male: 2.61 mSv
 
Female: 3.32 mSv
Contrast medium IV injection
Start delay
50 – 60 sec.
Flow rate
1.0 – 2.0 ml/sec.
Total amount
100 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 115  Friday, April 8, 2005  9:55 AM